Dialog box of "Download Recorded Data"

Download the recording data of the main unit of the Thermo Recorder.

Data display channel
The recording data of the main unit of the Thermo Recorder is divided into 2 channels. Channel registering sites of 1 to 8 data display channels for each channel is set up.
An empty channel is displayed for an initial value. However, setting up a channel where already recorded data is registered is available. In this case, the old data is cleared and the recorded data newly downloaded is registered.

Execute downloading
Click on the download button, and the downloading of recorded data starts.

Downloading conditions
When downloading starts, the Thermo Recorder unit model, channel name, record starting date and time, duration, recording conditions, recording mode and starting mode are displayed and the downloaded data volume is also displayed in an indicator.

Precautions
In some cases, downloading is unavailable depending on the condition of the main unit of the Thermo Recorder.


a) When the main unit of the Thermo Recorder has just been turned on, an error is shown because of no recorded data. During record programming with the main unit of the Thermo Recorder, an error is shown because of improper Thermo Recorder mode.

b) Communication is unavailable.
An unconnected communication cable or mistake in selecting a serial port causes an error.

c) Data can be downloaded even during recording.
Download the recording data from the start to the present. The main unit of the Thermo Recorder continues recording during and after downloading.

d) Save the recording data downloaded in a file as soon as possible.
The recording data of the main unit of the Thermo Recorder will be lost by turning off the power or newly starting recording. Further, the downloaded recording data will be lost by terminating this software or turning off the power of the computer. It is advised to save it in a file as soon as possible.