When using the adjustment function, measurements are displayed and recorded after being adjusted according to the conditions set by the user.
In the [Settings] menu of this software, select [Adjustment Settings] to open the setting window.
Ch.1, Ch.2 [Adjust this Channel] |
Place a check mark next to [Adjust this Channel] for the channel(s) to which you wish to make adjustments, select the adjustment method, and make the desired entries. (See: Adjustment Entries and Guidelines)
|
---|---|
Settings History | The box in the bottom shows a log for adjustment settings history.
|
[Close] Button | This quits and closes adjustment settings. |
[Send Settings] Button | This sends the settings to the connected device.
|
[Initialize] Button | This will reset the adjustment settings stored in the device/sensor (No setting: y=1x+0). |
[Save File] Button | This saves entries to a file. |
[Read File] Button | This reads entries which were saved to a file. |
Temperature, Humidity | ||
Slope (a) | Range | 0<a≤2 |
---|---|---|
Significant Digits | 3 decimal places | |
Intercept (b) | Range | within ±30 °C within ±30 % |
Significant Digits | same as that of the data logger's measurement resolution |
There are two possible methods of adjustment: "1 point Adjustment” and "2 point Adjustment”.
1 Point Adjustment | Use this to adjust all measurements with the same offset. (The whole graph will be adjusted up or down, keeping the same slope.) |
---|---|
2 Point Adjustment | Use this to adjust measurements in a wide range or when adjustment cannot be carried out using only one point. The slope will change. |
This adjustment method changes only the offset (b) where the slope (a) is considered 1.
For example, the data logger is measuring 10.2 °C, but the standard says the actual measurements should be 10 °C.
Enter 10.2 in the "Before Adjustment" box and 10 in the "After Adjustment" box.
The conversion equation will be Y=X−0.2 and for all measurements an adjustment to the offset of −0.2 will be made.
The adjustment equation will be calculated from two points: the slope (a) and the intercept (b).
For example, the data logger is measuring 0.2 °C and 9.6 °C, but the standard says the actual measurements should be 0 °C and 10 °C.
Enter 0.2 and 9.6 in the "Before Adjustment" boxes and 0 and 10 in the "After Adjustment" boxes.
In this case the slope (a) is (10−0)/(9.6−0.2)=1.06.
The conversion equation is Y=1.06X−0.213 with the intercept being −0.213.