By using Auto-upload and Warning Settings, it is possible to monitor for warnings on the server (T&D WebStorage Service / T&D Data Server), and to receive warning notifications by email.
Also, the [ALM] icon on the TR7A display notifies the user of the occurrence of a warning.
Warning notification methods and conditions for warnings/recovery may differ depending on the device model and data destination server. Please refer to [Specifications for Warning Transmission] for details.
Only checked items will be monitored for warning.
[Send Settings] Button | This will send the settings to the connected device. |
---|---|
[Get Settings] Button | This will retrieve the current settings from the connected device. This can be used to apply the same settings to multiple devices as shown below:
|
Upper Limit / Lower Limit | The checked items will be monitored for warning. Enter numeric values for each channel.
|
Sensor Warning | Select "ON" in order to monitor for sensor errors (No sensor connected, improper connection, damaged, etc.)
|
Warning Judgment Time | When an out-of-limit condition or sensor error continues for the set length of time, a warning will be judged to have occurred.
|
Battery Warning | Select "ON" if you wish to receive a warning notification when batteries need to be replaced. (TR-7wf is not supported)
|
*1: For how to check your firmware version, refer to [Q&A about TR7 Series Devices] - [Q12: How can I check my firmware version and update it when necessary?] *2: For the estimated battery life, refer to [Q&A about TR7 Series Devices] - [Q13: What is the battery life for the data loggers?] |
For the TR7A with upper/lower limit and/or sensor warning checked in the Warning Settings, the [ALM] icon will flash on the LCD screen when a warning is detected.
Note, however, that when the TR71A/75A in Vaccine Mode has an Upper or Lower Limit Warning, the [ALM] icon will not disappear until you clear the display.
When a measurement cannot be performed due to disconnection or malfunction of the sensor, the readings on the LCD will be shown as [--.-]. If the sensor warning setting has been set to ON, [ALM] will also flash.
In TR71A, even if the external sensor connected to Ch1 fails, the internal sensor for Ch1 will continue to measure uninterrupted and no error will be shown.
EX: Sensor Error in Ch2 (TR71A)
Use the DISPLAY button to switch the display to check which limit (upper or lower) has been exceeded.
(Refer to LCD Display Pattern in [Note]).
EX: TR71A/75A Ch2
Switch to the fixed display of Ch2 MAX/MIN values and check the position of the [ALM] icon.
If it appears next to [Max], it is an upper limit warning. If next to [Min], then it is a lower limit warning.
If you start a new recording session or change the warning settings while the device is in a warning state, the ALM display will disappear and you will receive a notification (Warning status: Warning Lifted).
The following explains how to turn off the display manually while the recording continues. After the operation, you will receive a notification (Warning status: Back to Normal).
Connect the device to your PC via USB, and go to [Communication] Menu - [Clear ALM Display].
In addition to using the "TR7 for Windows" software and the DISPLAY Button on the device, you can also use the "T&D Thermo" app to clear the ALM display manually. (For details, refer to the Help of T&D Thermo)
LCD Display Pattern
Ch1/Ch2 current values (fixed) → Ch1 MAX/MIN values (fixed) → Ch2 MAX/MIN values (fixed) → alternating display (for 2 seconds each)
The factory default setting is a fixed display of Ch1/Ch2 current values.