Making Warning Settings

By using Auto-upload and Warning Settings, it is possible to monitor for warnings on the server (T&D WebStorage Service / T&D Data Server), and to receive warning notifications by email.

Also, the [ALM] icon on the TR7A display notifies the user of the occurrence of a warning.

Note

Warning notification methods and conditions for warnings/recovery may differ depending on the device model and data destination server. Please refer to [Specifications for Warning Transmission] for details.

[Warning Settings] Tab

Only checked items will be monitored for warning.

Note

  • When there are multiple device icons in the Connected Devices box, select the one for which you wish to make settings.
  • Warning settings can also be changed from T&D WebStorage Service or T&D Data Server.

Go to Top

[ALM] Icon Display Examples (TR7A)

For the TR7A with upper/lower limit and/or sensor warning checked in the Warning Settings, the [ALM] icon will flash on the LCD screen when a warning is detected.

Note:

Note, however, that when the TR71A/75A in Vaccine Mode has an Upper or Lower Limit Warning, the [ALM] icon will not disappear until you clear the display.

Upper / Lower Limit Warning

Sensor Warning

When a measurement cannot be performed due to disconnection or malfunction of the sensor, the readings on the LCD will be shown as [--.-]. If the sensor warning setting has been set to ON, [ALM] will also flash.
In TR71A, even if the external sensor connected to Ch1 fails, the internal sensor for Ch1 will continue to measure uninterrupted and no error will be shown.

EX: Sensor Error in Ch2 (TR71A)

Upper / Lower Limit Warning Details

Use the DISPLAY button to switch the display to check which limit (upper or lower) has been exceeded.
(Refer to LCD Display Pattern in [Note]).

EX: TR71A/75A Ch2

Switch to the fixed display of Ch2 MAX/MIN values and check the position of the [ALM] icon.
If it appears next to [Max], it is an upper limit warning. If next to [Min], then it is a lower limit warning.

Go to Top

Clearing the ALM Icon (TR7A)

If you start a new recording session or change the warning settings while the device is in a warning state, the ALM display will disappear and you will receive a notification (Warning status: Warning Lifted).

The following explains how to turn off the display manually while the recording continues. After the operation, you will receive a notification (Warning status: Back to Normal).

Using the "TR7 for Windows" Software:

Connect the device to your PC via USB, and go to [Communication] Menu - [Clear ALM Display].

Using the DISPLAY Button on the Device:

  1. With each pressing of the DISPLAY button, the item on the display will change.
    When the desired display pattern (ALM/Max/Min) appears, stop pressing the button.
  2. Press-and-hold the DISPLAY button until the ALM icon disappears.
    The Max/Min values will also be cleared.
  3. Return to the original display by pressing the DISPLAY button again.

In addition to using the "TR7 for Windows" software and the DISPLAY Button on the device, you can also use the "T&D Thermo" app to clear the ALM display manually. (For details, refer to the Help of T&D Thermo)

Note:

LCD Display Pattern
Ch1/Ch2 current values (fixed) → Ch1 MAX/MIN values (fixed) → Ch2 MAX/MIN values (fixed) → alternating display (for 2 seconds each)
The factory default setting is a fixed display of Ch1/Ch2 current values.